Experimental investigations of a manually versus an electrically driven skull drill for bedside usage

  • \(\bf Background\) Manual skull drilling is an old but in modern neurosurgery still established procedure which can be applied quickly and universally in emergency situations. Electrical drilling requires more complex equipment and is usually reserved to the Operating Room (OR). It also seems desirable to apply an electrical drill for bedside usage but a suitable product does not exist so far. \(\bf Method\) Our experimental study using a manually and an electrically driven skull drill included a total of 40 holes drilled into synthetic biomechanical sheets. Half of the holes were produced with a prototype electrical drilling machine of the company Kaiser Technology and half of them with a traditional manual drill. Different drilling parameters such as the geometry of the borehole, the drilling forces and the drilling vibrations were captured during all experiments. \(\bf Results\) The electrical drilling needed higher vertical force by the operators and a longer time to penetrate the sheet. A reason was the relatively lower rotational speed provided by this particular drill. When drilling electrically the vibrations were substantially less which in turn led to a more precise shape of the holes (revealed by observation via a microscope). \(\bf Conclusions\) The electrification of bedside drilling can in principle enable emergency craniostomies to be performed with greater ease and accuracy. The power of the electric drive, however, must be at least equivalent to the power of the traditional manual drill. Otherwise, the vertical forces exerted on the scull by the operator become inhibitive. The challenge is to combine cost-efficiency and re-sterilizability of an electrically driven drilling machine which at the same time is small and simple enough to qualify for emergency applications.

Download full text files

Export metadata

Additional Services

Share in Twitter Search Google Scholar
Metadaten
Author:Anne CarolusGND, Wolfgang RichterGND, Claus-Peter FritzenGND, Kirsten SchmiederGND, Christopher BrenkeGND
URN:urn:nbn:de:hbz:294-71899
DOI:https://doi.org/10.1371/journal.pone.0215171
Parent Title (English):PLoS one
Publisher:Public Library of Science
Place of publication:San Francisco
Document Type:Article
Language:English
Date of Publication (online):2020/05/27
Date of first Publication:2019/04/18
Publishing Institution:Ruhr-Universität Bochum, Universitätsbibliothek
Tag:Open Access Fonds
data acquisition; neurosurgery; prototypes; skull; vibration
Volume:14
Issue:4
First Page:e0215171-1
Last Page:e0215171-11
Note:
Article Processing Charge funded by the Deutsche Forschungsgemeinschaft (DFG) and the Open Access Publication Fund of Ruhr-Universität Bochum.
Institutes/Facilities:Knappschaftskrankenhaus Bochum, Klinik für Neurochirurgie
Dewey Decimal Classification:Technik, Medizin, angewandte Wissenschaften / Medizin, Gesundheit
open_access (DINI-Set):open_access